Jurnal Mipa
Penentuan Ion Logam Aluminium dalam Sediaan Deodoran dengan Metode Spektrofotometer Serapan Atom (SSA)
FMIPA Unmul
Email : dinda.anno92@gmail.com
A research about the aluminium content in deodorant with AAS method has been carried out. To obtain the optimum measurement results, should be done with analytical performance parameters. The result showed good performance, indicated of the LoD parameters was given 0,30 mg/L, %CV was given 2,63% and %Recovery for the samples deodorant X = 115%, deodorant Y = 96% and deodorant Z = 90,7%. Based on the results, this method was applied to aluminium contained in deodorant brand X, Y and Z. method respectively 0,61 µg/g for deodorant X : 0,73µg/g for deodorant Y and 0,29 µg/g deodorant Z. The result can be concluded for this research sholved the AAS method.
Kata Kunci : deodorant, aluminium, AAS